A new approach for the detection of a nonfluorescent compound by CE-resonance Raman spectroscopy based on the sweeping-MEKC mode

Citation:
Tsai, CH, Chan PH, Lin CH, Chang TC, Chia CT.  2006.  A new approach for the detection of a nonfluorescent compound by CE-resonance Raman spectroscopy based on the sweeping-MEKC mode, Dec. Electrophoresis. 27:4688-4693., Number 23

Abstract:

A CE-resonance Raman spectroscopy (CE-RRS) method based on MEKC and sweeping-MEKC modes is described. A nonfluorescent compound, malachite green (MG), and a doubled Nd:YAG laser (532 nm, 300 mW) were selected as model compound and light source, respectively. In order to carry out a quantitative analysis of MG, a monochromator (effective bandwidth, 0.4 nm) was used to collect the specific Raman line at 1616 cm(-1) (N-phi and C-C stretch, corresponding to 582 nm when the wavelength of the exciting source was 532 nm). As a result, the LOD for MG was 10 ppm, based on the MEKC/RRS mode. This could be improved to 5 ppb when the sweeping-MEKC/RRS mode was applied. Furthermore, with the addition of nano-size silver colloids to the CE buffer the detection limits can be further improved, but the data obtained with surface-enhanced resonance Raman spectroscopy (SERRS) are less useful for quantitative purposes.